Calibration Commands 
  Copyright © Itech Electronic Co., Ltd.  32 
Chapter11  Calibration Commands 
CALibration:SECure:[STATe] 
Set protection mode enable or disable when calibrating the power supply. 
Order syntax 
CALibration:SECure:[STATe] {<ON|OFF>[<password>]} 
Parameters 
0|1|ON|OFF 'XXXX’ 
Example 
CAL:SEC 1 'XXXX’; //XXXX is the instrument model 
CAL:SEC OFF 
Query syntax 
CALibration:SECure:STATe? 
CALibration:INITital 
This order is used to reset the calibration parameters to default setting. 
Order syntax 
CALibration:INITital 
Parameters 
None 
CALibration:SAVe 
This order is used to save the calibration parameters in a nonvolatile memory. 
Order syntax 
CALibration:SAVe 
Parameters 
None 
CALibration:VOLTage:LEVel 
This  order  can  set  voltage  calibration  point.  P1,  P2,  P3  and  P4  must  be 
calibrated orderly. 
Order syntax 
CALibration:VOLTage:LEVel <point>