16
Represent power by 4 bytes of Hex. Lower bytes are in the front location, higher bytes are in
the later location. For example :power is 200.000W,Hex is 0X00030d40,4th byte is 0X40,
5th byte is 0X0d,6th byte is 0X03,7th byte is 0X00。
10. Set / Read resistance value CR mode(30H/31H)
1
st
byte
Start bit ( AAH )
2
nd
byte
Address (0—31,0XFF)
3
rd
byte
Command(30H/31H)
4
th
byte
The lowest byte of resistance value.
5
th
byte
The lower byte of resistance value.
6
th
byte
The higher byte of resistance value.
7
th
byte
The highest byte of resistance value.
8
th
to 25
th
byte
System reserve
26
th
byte
Sum code
Represent resistance value by 4 bytes of Hex. Lower bytes are
in the front location, higher bytes are in the later location. If resistance value is 200.000R,Hex
code is 0X00030d40,4TH byte is 0X40,5TH byte is 0X0d,6th byte is 0X03,7th byte is 0X00。
11. Set / Read CC mode transient current and timer parameter. (32H/33H)
1
st
byte
Start bit ( AAH )
2
nd
byte
Address (0—31,0XFF)
3
rd
byte
Command(32H/33H)
From 4
th
byte to 7
th
byte
Setting value of current A (Lower bytes are in the front location, higher bytes
are in the later location.)
From 8
th
byte to 9
th
byte.
Time value of timer A ((Lower bytes are in the front location, higher bytes are in
the later location) (1 represent 0.1mS)
From 10
th
to 13
th
byte
Setting value of current B (Lower bytes are in the front location, higher bytes
are in the later location)
From 14
th
to 15
th
byte
Time value of timer B (Lower bytes are in the front location, higher bytes are in
the later location) (1 represent 0.1mS)
16
th
byte
Transition operation mode (0 is CONTINUES, 1 is PULSE, 2 is TOGGLED)
From 17
th
to 25
th
byte
System reserve
26
th
byte
Sum code
12. Set / Read CV transient voltage and timer parameter.(34H/35H)
1
st
byte
Start bit ( AAH )
2
nd
byte
Address (0—31,0XFF)
3
rd
byte
Command(34H/35H)
From 4
th
to 7
th
byte.
Setting value of voltage A (Lower bytes are in the front location, higher bytes
are in the later location)
From 8
th
to 9th byte.
Time value of timer A (Lower bytes are in the front location, higher bytes are in
the later location) (1represent 0.1mS)
NOTE