SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 87
SOUR:RES:RANGE MIN
Query Syntax
[SOURce:]RESistance:RANGe? [ MINimum | MAXimum | DEFault ]
Returned Parameters
<NR3>
[SOURce:]RESistance:HIGH
[SOURce:]RESistance:LOW
This command is used to set the voltage high and low limit determind when the
load is in constant resistance.
Command Syntax
[SOURce:]RESistance:HIGH <NRf+>
[SOURce:]RESistance:LOW <NRf+>
Parameters
MINimum through MAX | MINimum | MAXimum | DEFault
Unit
V (volts)
*RST Value
MAXimum
MINimum
Examples
RES:HIGH 5
Query Syntax
[SOURce:]RESistance:HIGH? [MINimum|MAXimum|DEFault ]
[SOURce:]RESistance:LOW? [MINimum|MAXimum|DEFault ]
Returned Parameters
<NR3>