Calibration Commands
Copyright © ITECH Electronic Co., Ltd. 63
CALibrate:VOLTage[:LEVel]
This command is only effective in calibration mode. According to the reading of
external measuring meter to input the calibration voltage. Please select a
calibration grade before you input a calibration value(using the command
CALibrate:VOLTage:POIN).These constant do not exist in the nonvolatile
memory unless you have sent the command CALibrate:SAVE.
Command Syntax CALibrate:VOLTage[:LEVel] <NRf>
Parameters <external reading>
Unit V (volts)
Examples CAL:VOLT 3.2223
Related Commands CAL:STAT CAL:SAV
CALibrate:VOLTage:METEr:POINt
This command is only effective in calibration mode, is used to set the calibration
point in CV mode.P1, P2 are used in low voltage range, P3, P4 are used in high
voltage range. Only be used to calibrate voltage source and voltmeter.
Command Syntax CALibrate:VOLTage:METEr:POINt <point>
Parameters P1 | P2 | P3 | P4
Examples CAL:VOLT:METEr:POIN P2
Related Commands CAL:STAT CAL:SAV
CALibrate:VOLTage:METEr[:LEVel]
This command only be effective in calibration mode. According to the reading of
external measuring meter to input the calibration voltage. Before you input the
value, you should firstly select a calibration grade(with command
CALibrate:VOLTage:POINt).All these constant do not exist in the nonvolatile
memory unless you have send the command CALibrate:SAVE to save them.
Command Syntax CALibrate:VOLTage:METEr[:LEVel] <NRf>
Parameters <external reading>
Unit V (volts)
Examples CAL:CURR 3.2223
Related Commands CAL:STAT CAL:SAV