17
APPENDIX B:
Hardware Specications
ITEM DESCRIPTION
Monitor 19” monitor
Scanner
Scanner emits red laser light (680nm Class 1) as well as
white LED emissions.
Wireless LAN
LAN card provides local network communications with
wireless connectivity.
Mains Fuses
T3.15AL (3.15A, slow-blow) 250V glass tube fuses (5 x 20
mm).
Operating Power 100-240VAC- 50/60 Hz – 350VA (max)
Operating Temperature 18°to 26°C / 64.4° to 78.8°F
Storage/Transportation
Temperature
-5°to 50°C 23° to 122°F
Operating Pressure &
Altitude
Pressure: 520 mmHg to 760 mmHg (69.3 kPa to 101.3 kPa)
Altitude: 0 feet to 10,000 feet
Storage/Transportation
Pressure & Altitude
Pressure: 430 mmHg to 760 mmHg (57.3 kPa to 101.3 kPa)
Altitude: 0 feet to 15,000 feet
Relative Humidity Operating: 40% to 70%; Storage: 30% to 90%
Dimensions
Base Unit:
Height: 389 mm (~15 Inches)
Width: 459 mm (~18 Inches)
Depth: 123 mm (~5 Inches
Scanning Unit:
Length: 338.5 mm (~13 Inches)
Width: 53.5 mm (~2 Inches)
Depth: 69.8 mm (~3 Inches)
Net Weight
Base Unit: 11 kg (~24 lbs)
Scanning Unit: 0.47 kg (~1 lbs)
Shipping Weight