Turn on the power of an ionizer. Press PLATE CONTROL button red . The②()
measurements start automatically in order of positive decay time, negative decay
time and ion balance.
() () () () ()abcde
() ( ) ( ) ()fg h i
() ( ) () ( )jklm
a Press PLATE CONTROL button red when auto mode is displayed.() ( )
()Positive decay time measurement starts A LED in +DECAY button lights
b The plate is grounded and digital zero activation. GND lamp lights.()
c Positive high voltage is applied to the sensor plate. HV OUT lamp lights.()
d Internal timer ON when the plate voltage reaches +1,000 V.()
e Internal timer OFF when the plate voltage reaches +100 V or setting voltage.() ( )
f The plate is grounded and digital zero activation. GND lamp lights.()
()Negative decay time measurement starts A LED in -DECAY button lights
g Negative high voltage is applied to the sensor plate. HV OUT lamp lights.()
h Internal timer ON when the plate voltage reaches -1,000 V.()
i Internal timer OFF when the plate voltage reaches -100 V or setting voltage.() ( )
j The plate is grounded and digital zero activation. GND lamp lights.()
()Ion balance measurement starts A LED in IB button lights
k During ion balance measurement.()
l The end of ion balance measurement. The LED in IB button lights off. The plate is()
grounded and digital zero activation. GND lamp lights.
m The end of all measurements. Buzzer beeps once.()
Flow of LCD display