INSTRUCTION MANUAL CODE 80416 REV D 12/2023 Page: 32
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
2.3 Safety parts and functions
Temperature Protection:
- Temperature Protector (Steam Generator) Cut o current when the steam generator temperature is too high.
- Temperature Protector (Heating Ring) Cut o current when the heating ring temperature is too high.
Electronic Protection:
- Double Fuse It will be broken circuit if the connected power is too high or unstable.
- Electronic Filter It will lter the electromagnetic interference during working.
Mechanical Protection:
- Jiggle Switch To ensure the door is closed completely for avoiding the risk of safety.
- Tray Tong Avoid scald when removing articles from the chamber.
Control Part:
- Temperature Sensor (Inside Chamber) To measure temperature inside the chamber.
- Temperature Sensor (Heating Ring) To measure temperature of the heating ring.
- Temperature Sensor (Steam Generator) To measure temperature of the steam generator.
- Pressure Sensor To measure pressure of the chamber.
- PCB Control Control system for controlling all the process of sterilization.
CAUTION!! MANUFACTURER IS NOT HELD RESPONSIBLE FOR ANY ARBITRARY DISASSEMBLY, AMEND-
MENT WITH THE UNIT, BY UNAUTHORIZED PERSON OR UNPROFESSIONAL TECHNICIAN.
2.4 Operation risks
Please take attention to avoid the following operation risk during operation:
Scald Risk
- Every time open the door after nish the cycle, please keep an appropriate distance, because the chamber still have
rudimental steam with high temperature, avoid to be scalded.
- Every time open the door after nish cycle, please not touch the main door and chamber, because of high temperature,
and avoid to be scalded.
Pollution Risk
Please clean the chamber after every time use to avoid rudimental contamination left inside chamber.
2.5 Protective device
Plastic or cloth gloves are protective devices useful during loading and removing articles, to avoid scald.