8.3 Weekly cleaning
Cleaning Chamber, Trays and Rack
At least once a week, the trays and tray rack should be removed from the sterilizer chamber. The trays, tray rack and
chamber should be thoroughly cleaned to remove any deposits from the surfaces. Clean the trays, rack and chamber
(especially the bottom of the chamber) with appropriate antibiological cleaners. Wipe all residues from the surfaces with
a dampened, lint-free cloth.
WARNING
To prevent from collection of mineral deposits and corrosion of chamber components, use distilled or deion-
ized water only as specied. Clean chamber after each use if sterilizing saline solutions.
Cleaning water draining lter
Water draining lter might has been jammed by some dust because of use for a long-term, so eect of vacuum and drying
would be inuenced. Some tiny impurity might be deposited on the lter after a long-term use, blocking the lter, so as to
inuence the eect of the vacuuming and water discharging. The kinds of impurity come from smeary dust on the instru-
ments being sterilized or some calcication in the water. Keep cleaning of the inside chamber in order to make life-time of
lter much longer; please take the following advice for consideration:
- Use eligible distilled water;
- The instruments should be cleaned before placing in; it is good to use specied packing for the instruments with oil or
other impurity, don’t forget to seal up.
- Rotate the water lter which composed by lter net tube(A) and lter holder(B) inside the chamber. Cleaning the part A
and B, ensure that there is no any dirty thing on it.( We suggest you to clean by ultrasonic cleaning machine). Then set it
back and rotate to the hold bottom of the chamber.
8.4 Limpieza mensual
Limpieza del tanque de agua
There are some impurities and some toxins had been left behind in the reservoir because distilled water stored for a long
time, so you need to drain and clean regularly. According to the picture which shows below, loose the screw by screw-driver,
and open the cover to clean inside.
ATTENTION
BE SURE THAT USE DISTILLED WATER PROPERLY IN ORDER TO EXTEND THE STERILIZER LIFE-TIME.
DO NOT RAVE ABOUT THE STERILIZER WHEN THE TANK HAS BEEN FILLED.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the J.P. SELECTA AUTESTER ST DRY PV-B print 18L and is the answer not in the manual?

J.P. SELECTA AUTESTER ST DRY PV-B print 18L Specifications

General IconGeneral
BrandJ.P. SELECTA
ModelAUTESTER ST DRY PV-B print 18L
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals