EasyManua.ls Logo

Jackson Labs LC_XO - 10 Synchronization; Diagnostic Subsystem; Diagnostic:roscillator:efcontrol:relative; Diagnostic:roscillator:efcontrol:absolute

Default Icon
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LC_XO GPSDO User Manual
36 © 2012 Jackson Labs Technologies, Inc.
A health status of 0x0 indicates a properly locked, and warmed-up unit that is completely healthy.
4.5.10 SYNChronization?
This query returns the results of these six queries :
SYNChronization:SOURce:MODE?
SYNChronization:SOURce:STATE?
SYNChronization:LOCKed?
SYNChronization:HOLDover:DURation?
SYNChronization:FEEstimate?
SYNChronization:TINTerval?
SYNChronization:health?
4.6 DIAGnostic Subsystem
This subsystem regroups the queries related to the diagnostic of the OCXO.The list of the commands
supported for this subsystem is as follows:
DIAGnostic:ROSCillator:EFControl:RELative?
DIAGnostic:ROSCillator:EFControl:ABSolute?
4.6.1 DIAGnostic:ROSCillator:EFControl:RELative?
This query returns the Electronic Frequency Control (EFC) output value of the internal reference
oscillator. It returns a percentage value between -100% to +100%. :
4.6.2 DIAGnostic:ROSCillator:EFControl:ABSolute?
This query returns the Electronic Frequency Control (EFC) output value of the internal reference
oscillator. It returns a value in volts between 0 and 5 V
4.7 MEASURE Subsystem
The MEASURE commands are not supported on the LC_XO board.
4.7.1 MEASure:VOLTage?
This command is not supported on LC_XO, and will return undetermined values.
4.7.2 MEASure:CURRent?
This command is not supported on LC_XO, and will return undetermined values.