EasyManua.ls Logo

Jamr Technology B22 - Page 20

Jamr Technology B22
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
2)
* Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
3)
* Caution: This unit h as been thoroughly tested and inspe cted to assure proper
performance and operation!
4)
* Caution: this ma chine should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, this machine should be observed to verify
normal operation in the configuration in which it will be used
Guidance and manufacture’s declaration electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the user of the
device should assure that it is used in such an environment.
Emission test Compliance Electromagnetic e nvironment guidance
RF emissions
CISPR 11
Group 1
The
device use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The device is suitable for use in all establishments,
including domestic establishments other than domestic
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of device
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
Not applicable Mains power quality should be
that of a typical commercial or
hospital environment.