6
pulse rate can be safely used by patient. The routine clean and changing batteries can be
performed by the pa tie n t.
Warning:
Use of power adapters
1.Adapter: input 100-240v, 50/60hz output DC 5V 1A
2.Do not to position the device to make it difficult to operate the disconnection device while
using adaptor.
3.Do not be prone to water leakag e, high temperature, moisture, direct sunlight and more or
more corrosive gas environment. And Do not use this product in the above environment.
Caution: To avoid any possibility of accidental strangulation, keep this unit away from children
and do not drape tubing around your neck.
Caution: To avoid damaging the device, keep this unit away from children and pets.
Caution: The standard ma terial used for the bladder and tubing is latex-free.
Attention: Self-measurement means c ontrol, not diagnosis or treatment. Unusual values must
always be discussed with your doctor. Under no circumstances should you alter the dosages
of any drugs prescribed by your do ctor.
Attention: The pulse display is not suitable for checking the frequency of heart pacemakers!
Attention: In cases of irregular heartbeat, measurements made with this instrument s hould
only be evaluated after consultation with your doctor.
Note: To obtain th e gr eatest accuracy from your blood pressure instrument, it is recommended
that the instrument be used within the specified tem perature and the relative humidity, please
see the Technical Specifications
Note: The cuff is treated as the applied part.The user should contact the manufacturer for
assistance, if needed, in setting up, using or maintaining the device.
2. Importan t Information on Blood Pressure and its Measurement
2.1. How does high or low blood pressure arise?
Your level of blood pressure is dete rmined in the circulatory center of the brain and ad justs to a
variety of situations through feedback from the nervous system. To adjust blood pressure, the
strength and spe ed of the heart (Pulse), as well as the width of circulatory blood vessels is
altered.Blood vessel width is controlled by fine muscles in the blood vessel walls.
Your level of arterial b lood pressure changes periodically during heart activity: During the
“blood ejection” (Systole) the value is highest (systolic blood pressure value). At the end of the
heart’s “rest period” (Diastole) pressure is lowest (diastolic blood pressure value).
Blood pressure values must lie within certain norma l ranges in order to prevent partic ular
diseases.
2.2. Which values are normal?
Please refer to the diagram below(Picture-01)