8. Certifications
Device standard:
This device is manufactured to meet the European blood pressure monitors:
IEC 80601-2-30/ ISO81060-1/ IEC60601-1-11/ IEC60601-1
Electromagnetic compatibility:
Device fulfills the stipulations of the International standard
IEC60601-1-2
9. Technical Specifications
Model: B22
Wight: 260g (batteries and AC adapter is not included)
Display: 3.74”x2.52” LCD Digital Display
Size: 144(W) x 120 (L) x 67(H) mm
Accessories: 1×Main Device, 1×Cuff, 1×Users manual, 1×Warranty card
Operating Conditions: Temperature: 5 to 40 ;Hu midity: 15% to 93 % RH;
Pressure altitude: 70KPa~ 106Kpa
Storage And Shipping Conditions:Temperature: -25 to 70; Humidity: 93% RH;
Pressure altitude: 70KPa~ 106Kpa
Measuring method: Oscillometric
Operating method: Continuous op eration
Pressure sensor: Resistive
Measuring range:
DIA: 30-195mmHg; SYS: 60-250mmHg
Pulse: 40 to 199 per minute
Cuff pressure display range: 0–295 mmHg
Memory: Automatically stores the last 120 measurements for 2 users (total 240)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: a) 4 AAA batteries, 1.5 V
b) AC adapter INPUT 100-240VAC 50/60HZ OUTPUT5VDC 1A
Accessories: Wide range rigid cuff : 8.7” 15.7” (22 - 40 cm)
Automatically power off: 60 seconds
Users: Adult
Software Version: V1.1.1
Expected service life of the device and accessories: 5 years
Technical alterations reserved!
9.1.
EMC Declaration
1)
* This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be affected by
portable and mobile RF communications equipment.