P/N: 0302-0375B
MODEL PU-1580
Intelligent HPLC Pump
Operation Manual

Table of Contents

Other manuals for Jasco PU-1580

Questions and Answers

  • H
    Heather FowlerSep 4, 2025
    Why is the retention time shortening in my Jasco Water Pump?
    • C
      clarkmelodySep 4, 2025
      Shortening retention time in your Jasco Water Pump can be attributed to column deterioration; consider replacing the column. Variation in ambient temperature can also affect retention time; protect the column using thermal insulation. Another cause can be variation in eluent composition; prepare the eluent again.
  • K
    Kimberly RothAug 31, 2025
    Why is the retention time lengthening in my Jasco Water Pump?
    • D
      davisjerryAug 31, 2025
      Lengthening retention time in your Jasco Water Pump can occur due to a decrease in the actual flow rate; see Section 2.1.2 for details. Insufficient column re-equilibration can also cause this; increase the column re-equilibration time. Variations in ambient temperature can affect retention time; protect the column using thermal insulation. Changes in eluent composition can also contribute; prepare the eluent again.
  • S
    Stefanie DennisAug 30, 2025
    Why does Jasco Water Pump have spike-shaped baseline noise?
    • L
      Larry DawsonAug 30, 2025
      If you observe spike-shaped baseline noise in your Jasco Water Pump, it could be due to bubbles in the cell, which causes noise toward increasing absorption; refer to the maintenance manual for the UV detector. Another cause could be electrical noise; in this case, ground the instrument.
  • S
    Sara LiAug 28, 2025
    Why is the bottom panel drain port of my Jasco PU-1580 leaking?
    • B
      brandoncrawfordAug 28, 2025
      If the bottom panel drain port of your Jasco Water Pump is leaking, it could be due to deterioration of the diaphragm on the plunger washing mechanism; in this case, replace the diaphragm. Alternatively, it may be due to leakage from the plunger seal. If the head mount flange hole is clogged, solvent will leak from the drain hole.
  • C
    Christopher ZamoraAug 26, 2025
    Why does my Jasco PU-1580 Water Pump have high baseline noise?
    • Y
      ykimAug 27, 2025
      High baseline noise in your Jasco Water Pump can stem from several causes. Variations in pump pressure can contribute to baseline noise; see Section 2.1.3 for related information. Mixing noise, especially when running a gradient program or using a post-column derivatization system, can also cause this issue; consider adding an extra mixer or damper to the pump. Detector cell problems like bubbles, contamination, or leakage can also lead to high baseline noise; refer to the maintenance manual of the UV detector. A decrease in the light intensity of the detector, due to light source or mirror deterioration, can also cause this issue; refer to the maintenance manual of the UV detector.
  • M
    Megan LopezAug 21, 2025
    What to do if Jasco PU-1580 Water Pump shows TROUBLE UNDER PRESS errors?
    • A
      Aaron VasquezAug 21, 2025
      If the Jasco Water Pump experiences 'TROUBLE UNDER PRESS' errors, it might be due to air bubbles in the pump head. To resolve this, open the purge valve to remove the air bubbles. Another cause could be a clogged inlet filter, especially if using buffer solutions; in this case, replace the inlet filter. Also, check for a leaking plunger seal and tighten the pump head securing screws. If leaking continues, replace the plunger seal. A leaking inlet tube fitting can also cause this issue; tighten the fitting. In some cases, the pump head might be deteriorating, requiring thorough cleaning with an acid or alkaline solution. If the leak occurs at the downstream side of the pump, refer to the appropriate manual.
  • S
    Susan GonzalesDec 3, 2025
    What causes large baseline drift in Jasco Water Pump?

    Summary

    Safety Considerations

    Safety Symbols

    Explains the meaning of hazard warning symbols used in the manual.

    Warning Labels

    Describes the warning labels attached to the instrument and their significance.

    Limited Warranty

    Regulatory Statements

    CE Notice

    Details compliance with European Community directives and technical standards.

    Preface

    Installation Conditions

    Specifies environmental and operational requirements for safe and optimal instrument setup.

    Maintenance

    Product Description and Specifications

    Product Description

    Overview of the PU-1580 Intelligent HPLC Pump's features and capabilities.

    Specifications

    Detailed technical specifications and performance parameters of the HPLC pump.

    Part Names and Descriptions

    Front Panel

    Identifies and describes components on the front of the instrument.

    Operation Panel

    Details the keys and their functions on the operation panel.

    Rear Panel

    Describes connectors and ports located on the rear of the unit.

    Power-ON and Power-OFF

    Power-ON and Diagnostic Test

    Procedure for turning on the instrument and its automatic self-test sequence.

    Power-OFF

    Instructions for safely shutting down the instrument and saving settings.

    Operations in Normal Operation Mode

    Switching Between Constant Flow and Constant Pressure Modes

    How to change between constant flow and constant pressure delivery modes.

    Constant Flow Mode

    Detailed operation and parameter setting within constant flow mode.

    Constant Pressure Mode

    Detailed operation and parameter setting within constant pressure mode.

    Error Messages During Pumping

    Troubleshooting common error messages encountered during pumping operations.

    Operations in Program Mode

    Description

    Overview of program mode, gradient delivery, and time programs.

    Editing Time Programs

    Creating and modifying time-based program steps for gradients.

    Example of a High-Pressure Gradient Program

    Illustrates a typical high-pressure gradient program setup and execution.

    Low-Pressure Gradient Mode

    Details of operating in low-pressure gradient mode, including parameter setup.

    High- or Low-Pressure Gradient Operations

    High- or Low-Pressure Gradient Program Execution

    How to execute programmed gradient sequences.

    Parameter Changes While the Gradient Program is Running

    Modifying settings such as flow rate and pressure during gradient execution.

    Special Operations

    Timer Functions

    Setting automatic start (ON) and stop (OFF) times for pumping.

    Flow Rate Correction

    Adjusting flow rate accuracy and matching flow rates between pumps.

    Switching Between High- and Low-Pressure Gradient Modes

    Changing the gradient operation mode based on system requirements.

    Adjustment of LPG COMP [SHIFT][7]

    Fine-tuning LPG proportioning valves for accurate solvent composition.

    Using the Optional Semi-Micro Kit

    Activating special programs for low flow rate semi-micro applications.

    Jasco PU-1580 Specifications

    General IconGeneral
    BrandJasco
    ModelPU-1580
    CategoryWater Pump
    LanguageEnglish