Err 5 Light Saturation Not Found
This error indicates that the peak light hasn’t been found at zero.
The most likely causes of this error are:
Fatal 1. Lamp failure.
2. Deterioating lamp signal.
3. Sample or cuvette in the sample holder.
Solution: Ensure that the sample holder is empty. Restart the unit, if the
problem persists contact a service technican.
Err 7 USB Module Failure
This error indicates the instrument is unable to detect a response from
the UBS module.
Warning
The most likely causes of this error are:
1. USB port failure.
Solution: press the key adjacent to the tick icon to carry on using the
instrument. Restart the unit, if the problem persists contact a service
technician.
Err 8 Unable to find Vane on Turret
This error indicates that the vane position zero on the turret cannot
be found.
Warning
The most likely causes for this error are:
1. The turret carousel has been removed and not replaced.
Redo
Solution: Check if the turret is in the sample chamber and inserted
properly. Press the key adjacent to the back icon to try rechecking.
Err 9 Over Temperature
This error indicates that the thermal switch has cut out.
The most likely causes for this error are:
Fatal 1. Failure of the fan
2. Thermal switch not connected.
Solution: restart the unit, if the problem persists contact a service
technician.