Joerns® Support Surfaces
P.R.O. Matt® Plus
© 2019 Joerns Healthcare • 6110394 RevB • 19-2275
Filter Replacement
When using an optional control unit, check the air
lter on the side of the unit periodically for buildup of
dust/dirt. If buildup is visible, turn o󰀨 the control unit
and disconnect the power cord from the wall outlet. To
access the lter, rst remove the lter cover, and then
remove the lter. Replace with a new lter. Ensure the
lter covers the entire lter region after replacement.
Reinstall the lter cover.
Steam Cleaning
Do not use any steam cleaning device on the unit.
Excessive moisture can damage mechanisms in
this unit.
Do not use any steam cleaning device on the mattress.
Excessive moisture can damage mechanisms in this
mattress.
Control Unit Cleaning
Wipe o󰀨 dust. If necessary, clean the housing exterior
with a disinfectant solution or a mild detergent and a
damp cloth, then wipe dry. Wipe power cord.
Caution: Do not use phenol based cleaning solution.
Equipment damage could occur.
Periodic Inspection
Joerns Healthcare recommends the P.R.O. Matt
®
Plus control unit, lter, power cord, and hose set be
inspected at least once a year for proper operation.
Warning: Only authorized personnel trained by Joerns
Healthcare should perform preventative maintenance.
Preventative maintenance performed by unauthorized
personnel could result in personal injury and/or
equipment damage.
Any maintenance done without Joerns Healthcare’s
authorization will void any warranties on this product.
Storage and Care
Support Surface
Thoroughly wipe down outside of the support
surface as described in previous section and allow to
air dry prior to storage.
Cover with plastic and return to storage area. It is
recommended not to fold the mattress and to avoid
storage of the mattress other than in a at format.
Control Unit
Detach the hose set from the control unit and carefully
coil for storage with the unit.
The power cord may be wrapped around the unit for
convenience. Wrap the unit in a plastic bag for dust
resistance, then store the unit in an area appropriate
for an electronic medical device.
Note: When the product is not in use, store the power
cord properly. Failure to do so could result in personal
injury.
Waste Disposal
This product has been supplied from an
environmentally aware manufacturer that complies
with the WEEE. Please be environmentally responsible
and recycle this product through your recycling facility
at its end of life or dispose of it in accordance with
local regulations.