14
Joerns® Support Surfaces
P.R.O. Matt® Plus
© 2019 Joerns Healthcare • 6110394 RevB • 19-2275
Appendix A: Electromagnetic Compatibility (EMC) Related Notifications
Recommended separation distances between portable and mobile RF communications
equipment and the P.R.O. MATT Plus
The P.R.O. Matt Plus Control Unit (PMP-CU) is intended for use in an electromagnetic environment (for home
healthcare and professional healthcare) in which radiated RF disturbances are controlled. The customer or the
user of the P.R.O. Matt Plus Control Unit (PMP-CU) can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the P.R.O.
Matt Plus Control Unit (PMP-CU) as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=1,2√P
80 MHz to 800 MHz
d=1,2√P
800 MHz to 2,7 GHz
d=2,3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reection from structures, objects, and people.
Guidance and Manufacturer’s Declaration –Electromagnetic Emissions
The P.R.O. Matt Plus Control Unit (PMP-CU) is intended for use in the electromagnetic environment (for home
healthcare and professional healthcare) specied below. The customer or the user of the P.R.O. Matt Plus Control
Unit (PMP-CU) should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance (for home healthcare
and professional healthcare environment)
RF emissions
CISPR 11
Group 1
The P.R.O. Matt Plus Control Unit (PMP-CU) uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The P.R.O. Matt Plus Control Unit (PMP-CU) is suitable for use in all
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations /
icker emissions
IEC 61000-3-3
Compliance