Guidance and Manufacturers Declaration - Electromagnetic Immunity:
The P.R.O. Matt Plus Control Unit (PMP-CU) is intended for use in the electromagnetic environment (for home
healthcare and professional healthcare) specied below. The customer or the user of the P.R.O. Matt Plus Control
Unit (PMP-CU) should assure that it is used in such an environment.
Immunity Test IEC60601 Test Level Compliance Level Electromagnetic Environment-Guidance (for
home healthcare and professional healthcare
environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact: ±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Contact: ±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
Not applicable
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Surge
IEC 61000-4-5
±0.5kV,±1kV line(s)
to line(s)
±0.5kV,±1kV,±2kV
line(s) to earth
±0.5kV,±1kV line(s)
to line(s)
Not applicable
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
Voltage dips:
0% Uτ; 0,5 cycle
0% Uτ; 1 cycle
70% Uτ; 25/30 cycles
Voltage interruptions:
0% Uτ; 250/300
cycle
Voltage dips:
0% Uτ; 0,5 cycle
0% Uτ; 1 cycle
70% Uτ; 25/30 cycles
Voltage interruptions:
0% Uτ; 250/300
cycle
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment. If the user of the P.R.O. Matt Plus
Control Unit (PMP-CU) requires continued
operation during power mains interruptions, it is
recommended that the P.R.O. Matt Plus Control
Unit (PMP-CU) be powered from an uninterruptible
power supply or a battery.
Power
frequency (50,
60 Hz) magnetic
eld IEC 61000-
4-8
30 A / m
50 Hz or 60 Hz
30 A/m
50 Hz, 60 Hz
The P.R.O. Matt Plus power frequency magnetic
elds should be at levels characteristic of a
typical location in a typical home healthcare and
professional healthcare environment.
Note: Uτ is the a.c. mains voltage prior to application of the test level.
*During DIP interference, the pump will outage these are normal. The pump outage does not a󰀨ect the motor
operation.