Manufacturer’s declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The P.R.O. Matt Plus Control Unit (PMP-CU) is intended for use in the electromagnetic environment (for home
healthcare and professional healthcare) specied below. The customer or the user of the P.R.O. Matt Plus Control
Unit (PMP-CU) should assure that it is used in such an environment.
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
Compliance LEVEL
(V/m)
(for home healthcare
and professional
healthcare environment)
385 380-390 TETRA 400
Pulse
modulation b)
18 Hz
1,8 0,3 27 27
450 430-470
GMRS 460,
FRS 460
FM c)
±5 kHz
deviation
1 kHz sine
2 0,3 28 28
710
704-787
LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28
870
930
1720
1700-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2 0,3 28 28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2 0,3 28 28
5240
5100-
5800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 95500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
* Recommended separation distances between other equipment and this device - avoid stacking and locating it near
other electronic devices.
* If abnormal behavior is observed due to EM disturbances, please relocate the device accordingly. To avoid negative
inuence on electromagnetic compatibility, only use attachments and/or accessories that are recommended by the
manufacturer.