Menu functions
38
Warning:
The working temperature range of the Cryo-Compact Circulator is determined during
configuration. If set to >Control<, this range cannot be exceeded.
If set to > serial < and > eprog <, heat or cold is applied to the bath without control.
The permissible maximum temperature can be exceeded. The user has to take
adequate precautions for temperature control.
Materials, such as gaskets or insulations for example, may be damaged or destroyed, if
the permissible maximum temperature is exceeded.
The safety and warning functions of the instrument should always be used to their
fullest capacity. (See page 30)
Example:
Setpoints
>Setpoint1: 20.00°C
Setpoint2: 37.00°C
Sezpoint3: 70.00°C
°C °F
Setpoints
>Setpoint1: 68.00°F
Setpoint2: 98.60°F
Setpoint3: 158.00°F
Unit
Temperature values can be displayed in the unit °C or °F.
With this change all adjusted temperature values are converted and displayed
into the new unit.
-OFF-
S xxxx
I xxxx
°F
Possible parameters:
°C (factory setting)
°F