EasyManua.ls Logo

Julabo CF31 - Page 56

Julabo CF31
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Menu functions
56
Eprog : 50.0°C
IntAct : 24.7°C
Power : 80 %
Control : intern
After returning the LCD display to standard display by pressing
the temperature value adjusted and set on the external voltage or
current source is displayed in line 1
(Example: Eprog : 50.00 °C).
This EPROG input enables the use of different voltage and current
values as program parameters.
20
40
60
80
100
200
300
250
150
°C
1 2
3 4 5
6 7 8 9 10 V
2 4 6
8
10 14 18 20 mA
A
B
L Value“ - Setting the LOW value: (See below
)
1) Adjust and set the lowest desired value on the voltage or
current source resp. (Example A: 1 V).
Wait appr. 30 seconds.
2) Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the instrument (Example A: 20 °C ) and set by
pressing enter
.
„H Value“ - Setting the HIGH value: (See below
)
1) Adjust and set the highest desired value on the voltage or
current source resp. (Example A: 10 V).
Wait appr. 30 seconds.
2) Assign an upper temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the instrument
(Example A: 200 °C) and set by pressing enter
.
Return to the standard display by pressing
.
Example B in the diagram serves to illustrate that the end point
values are freely selectable (Ex: 8 mA and 16 mA).
>Value as: Setpoint
Signal : Voltage
L Value : 20.0°C
H Value : 200.0°C
Setpoint: 152.0°C
Input EPROG
Eprog : 152.00°C
IntAct : 50.76°C
Power : 100 %
Control : intern
Example out of diagram A:
Adjusting the voltage source for an output of 7.6 V!
Line 5 of the LCD DIALOG-DISPLAY shows the externally set setpoint
value. The instrument calculates this value from the rise angle of the two
pre-decided end points (in example A: 7.6 V correspond to an external
setpoint temperature of 152.0 °C ).
After returning the LCD display to standard display by pressing
,
this value is displayed in line 1
(Example: EPROG 152.00 °C).
Notice:
If this adjustment is not correctly performed at two different points, the setpoint setting
will be incorrect.
Important:
The usable temperature range between > L Value < and > H Value < is limited to the
configured working temperature range of the Cryo-Compact Circulator resp. of the unit
combination (working temperature range see technical specifications on page 11)

Table of Contents

Related product manuals