Appendix
66
2.1.2 OUT commands
OUT commands set device parameters. Remote control mode must be active.
Parameter
settings
Parameter Setting
out_sp_00
xxx.xx Setting for the setpoint temperature
out_sp_03
xxx.xx Setting for the high temperature
out_sp_04
xxx.xx Setting of the low temperature
out_sp_07*
x
Set pump to preset levels 1…5
out_sp_27
xxx Pump capacity in %
* Do not use for reprogramming, use out_sp_27 instead. Pump stage 4 has an output pressure of
approx. 450 mbar.
Device modes Parameter Setting
out_mode_05
x Start/stop command of the device in remote control
mode:
0 = Stop tempering
1 = Start tempering
2.1.3 Status commands
Status commands are used to query the current status of the device.
Status commands System response
version
Current firmware version
status
Return of status, error, warning, alarm