EasyManua.ls Logo

Julabo FK30-SL - List of Commands

Julabo FK30-SL
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control
84
Important times for a command transmission:
To ensure secure data transfer, the time gap between two commands
should be at least 250 ms.
The circulator automatically responds to an IN command with a data string
followed by an LF (Line Feed). The next command should be sent only
after 10 ms.
The OUT commands are valid only in remote control mode.
Examples:
When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A032)
Command to set the working temperature >Setpoint1< to 55.5 °C
OUT_SP_00
⇔
55.5
↵
A032_OUT_SP_00
⇔
55.5
↵
Command to retrieve the working temperature >Setpoint1<
IN_SP_00
↵
A032_IN_SP_00
↵
Response from the circulator:
55.5
↵
LF
A032_55.5
↵
LF
Notice: OUT commands
Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.
12.3. List of commands
OUT commands: Setting parameters or temperature values.
Command Parameter Response of circulator
OUT_MODE_01
0 Use working temperature >Setpoint1<
OUT_MODE_01
1
Use working temperature >Setpoint2<
OUT_MODE_01
2 Use working temperature >Setpoint3<
OUT_MODE_02
0 Selftuning “off”.
Temperature control using the stored parameters.
OUT_MODE_02
1 Selftuning “once”
Single selftuning of the controlled system after the next start.
OUT_MODE_02
2 Selftuning “always”
Continual selftuning of the controlled system whenever a new
setpoint is to be reached.

Table of Contents