1. Clean the probe
45$.
459$$$ (
$?
4856(..$$
?
2. High-level disinfection
#$
45($. 7
45 T 9
$
R>%
2P 9P &
R4:M5
2!R
#
P>>%
2P 9P &
#
4DDM5
2!;#6
#
Please follow the instructions about disinfectant concentration and disinfection method,
as well as the precautions about disinfectants provided by disinfectant provider. But do
not rinse the cable close to probe connector.
The soaking time of probe in the disinfectant is limited to the minimum time
recommended by disinfectant manufacturer (e.g., Cidex OPA manufacturer
recommended minimum 12 minutes).
Please follow local laws and regulations to choice the disinfectants.
4856($$45
T$
4:56(..$$
?
Attention:
1. In the process of cleaning and disinfection, avoid probe overheat (exceeding 60
℃
) as it
may be deformed or damaged under excessive heat.
2. In the operation of disinfection, please refer to medical institutions disinfection
technical specifications.
Attention:The waterproof grade of probe is IPX4, immersion depth as shown below.
Below 5mm metal ring can be immersed in liquid.