7.
troubleshooting
guide
Listed
below
are
some
of
the
probable
causes
of
alarm
conditions
and
their
corrections.
Symptom
;
Probable
Cause
|
Corrections
No
Set
Pump
set
has
Check
to
make
sure
that
only
Kangaroo
Pump
Set
Alarm
been
improperly
|
has
been
placed
onto
the
pump.
(no
SEt)
|
placed
onto
Check
to
make
sure
that
the
pump
set
has
been
pump.
properly
positioned
on
the
pump.
Flow
Error
|
Feeding
Refill
feeding
container.
if
feeding
set
has
been
Alarm
container
is
in
use
for
24
hours,
replace.
(FLO
Err)
|
empty.
Feeding
tube
or
|
Locate
point
of
occlusion
and
correct.
Note:
occlusion
feeding
set
may
occur
upstream
or
downstream
from
pump.
tubing
is
©
Feeding
Container
occluded.
*
Feeding
Pump
Set
(control
clamp
closed)
*
Feeding
Tube—feeding
tube
should
be
checked
for
obstruction.
Drip
chamber
Check
to
make
sure
formula
is
not
preventing
drop
walls
are
coated
|
detector
from
proper
operation.
if
formula
cannot
with
feeding
be
removed
from
inside
walls
of
chamber
via
formula.
chamber
manipulation,
replace feeding
set.
Sensor
in
upper
|
Check
to
make
sure
black
surface
of
drip
chamber
drip
chamber
guide
is
free
of
dried
formula.
Remove
any
deposits
guide
is
blocked.
|
by
using
a
cotton
swab
dampened
with
warm
soapy
water
or
isopropyl
alcohol.
Low
Battery has
been
|
Insert
pump
into
charger
base
and
plug
power
cord
Battery
run
down.
into
wall
outlet.
Alarm
(LO
bAt)
Hold
Error
|
Pump
has
been
|
Press
START/HOLD
to
silence
alarm,
then
press
Alarm
left
in
the
hold
START/HOLD
again
to
restore
flashing
rate
display,
(HLd
Err)
|
mode
for
over
and
press
START
/HOLD
again
to
restart
pump.
25
minutes.
System
Problem
with
Return
pump
for
technical
service.
Error
rotor,
rotor
Alarm
position
sensing
(SYS
Err)
|
or
other
rotor
circuit
problem.
Internal
Problem
with
Return
pump
for
technical
service.
Circuit
internal
pump
-
Problem
circuitry.
Alarm
(8888)
Dose
Preset
dose
has
|
Repeat
the
dose
amount
by
clearing
the
volume
Delivered
|
been
delivered
delivered;
OR
cancel
the
dose
feature
by
pressing
Alarm
to
patient.
DOSE
and
CLR;
OR
to
feed
additional
volume,
(dOSE
increase
the
dose
amount
by
pressing
DOSE,
then
dEL)
A
until
new
dose
is
displayed;
OR
turn
the
pump
off.
25