2.
maintenance
(cont.)
recommended
performance
test
SERIAL
NUMBER:
Date:
EQUIPMENT
CONTROLNO:
TECHNICIAN:
1.
Exterior
Cleaning/Inspection
a.
Power
cord
not
damaged
b.
Rotor
rollers
turn
freely
c.
Clean
case,
rotor,
and
drop
sensor
window
(refer
to
Service
Manual)
d.
Check
rotor
shaft
for
straightness
2.
Touch
Panel
and
Display
a.
“ON”
button
.
b.
“OFF”
button
c.
“A”
button
d.
“Y”
button
e.
“START/HOLD"
button
f.
“VOLUME”
button
g.
“DOSE”
button
h.
“INT”
button
i.
“CLR”
button
3.
Visual
Indicators
AC/Quick
charge
light
(on
charger
base)
NO
SET
Alarm
SYS
ERR
Alarm
FLO
ERR
Alarm
HLD
ERR
Alarm
.
Intermittent
Feeding
Indicator
(flashing
LED
on
pump
display)
-0
appa
4..
Pump
Motor
Cycle
Times
(with
feeding
set
in
place)
A.
Filla
feeding
set
with
water
and
connect
it
to
the
pump.
Do
not
place
the
pump
in
the
backpack/frame
at
this
time.
.
Turn
the
pump
ON
and
set
the
delivery
rate
to
10
mi/hr.
B
C.
Using
an
oscilloscope
or
timer/counter,
measure
the
time
period
from
the
beginning
of
one
pumping
cycle
to
the
next.
The
beginning
of
a
pumping
cycle
is
defined
by
the
rising
edge
of
voltage
being
applied
to
the
motor.
Record
the
cycle
time
below
and
compare
to
the
chart
in
the
Service
Manual.
D.
Repeat
step
C
at
a
delivery
rate
of
400
mi/hr.
E.
Repeat
steps
C
and
D
with
the
pump
in
the
backpack/frame.
a. In
Backpack/Frame
10
ml/hr
sec
400
mi/hr
sec
b..
Not
in
Backpack/Frame
10
ml/hr
sec
400
mi/hr
sec