7 Setup methods according to DIN EN ISO 17664
7.1 Preparation at the site of use
WARNING
Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
Observe suitable personal protective measures.
Remove all residual cement, composite or blood without delay.
Recondition the medical device as soon as possible after treatment.
Remove the cutter or grinder from the medical device.
The medical device must be dry when transported for reconditioning.
Do not place it in a solution or similar.
Setup methods according to DIN EN ISO 17664 35