5: Leakage current and insulation resistance Model 2470 High Voltage SourceMeter Instrument
User's Manual
5-6 2470-900-01 Rev. A / May 2019
View the measurements on the front-panel graph
To view the leakage current measurements on the front-panel graph:
1. Press the MENU ke
y.
2. Under Views, select Graph.
3. Select the Scale tab.
4. Under Y-Axis, next to Scale Format, select Log.
5. Under X-Axis, next to Method, select All.
6. Select the Graph tab to view the graph.
The following figure shows the front-panel graph for this application.
Figure 24: Front panel graph of leakage current measurement