7: TSP command reference  Series 2600B System SourceMeter® Instrument 
 
7-154  2600BS-901-01 Rev. B / May 2013 
 
When there are outstanding overlapped commands to acquire data, n refers to the index that the last entry in the 
table will have after all the readings have completed. 
If you pass a reading buffer instead of a reading buffer subtable, the default subtable for that reading buffer will 
be used. 
This command generates a single response message that contains all data. The response message is stored in 
the output queue. 
The format.data attribute controls the format of the response message. 
The following bufferVar attributes can be used with the print buffer command. 
bufferVar.dates, bufferVar.fillmode, bufferVar.formattedreadings, bufferVar.fractionalseonds, bufferVar.logstate, 
bufferVar.seconds, buffer.units, bufferVar.relativetimestamps, bufferVar.seconds, bufferVar.sourcestatuses, 
bufferVar, sourceunits, bufferVar.sourcevalues, bufferVar.statuses, bufferVar.times, bufferVar.timestamps, 
bufferVar.units 
If you are using the printbuffer() command to print multiple readings from multiple buffers, the readings must be 
paired: buffer 1 readings, buffer 2 readings, next buffer 1 readings, next buffer 2 readings and so on. For 
example, when you are storing voltage readings in one buffer and current readings in another buffer, keep them 
paired in the response. Or, when you are just storing different readings in multiple buffer keep then paired in 
response.