EasyManuals Logo
Home>Keithley>Measuring Instruments>2635

Keithley 2635 Reference Manual

Keithley 2635
562 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #314 background imageLoading...
Page #314 background image
12-54 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 12: Instrument Control Library Series 2600 System SourceMeter® Instruments Reference Manual
smuX.contact.calibratehi
X = SMU channel (a or b)
Function Calibrate the high/sense high contact check measurement.
Usage smuX.contact.calibratehi(cp1measured, cp1reference, cp2measured,
cp2reference)
cp1measured -- the value measured by this SMU for calibration
point 1.
cp1reference -- the reference measurement for calibration point 1
as measured externally.
cp2measured -- the value measured by this SMU for calibration
point 2.
cp2reference -- the reference measurement for calibration point 2
as measured externally.
Remarks • Contact check measurement calibration does not require range information.
• Typically the two calibration points used will be near 0Ω for calibration point 1 and 50Ω for
calibration point 2.
• All four measurements (cp1measured, cp1reference, cp2measured, and
cp2reference) must be made with the active calibration set. Corruption of the
calibration constants may result if this is not heeded.
• The new calibration constants will be activated immediately but they will not be
written to non-volatile storage. Use smuX.cal.save to commit the new constants to
nonvolatile storage. The active calibration constants will stay in effect until the
instrument is power cycled or a calibration set is loaded from non-volatile storage with the
smuX.cal.restore function.
• This function will be disabled until a successful call to smuX.cal.unlock is made.
Details See Section 16 "Calibration".
Also see smuX.contact.calibratelo
Note: 2635/36 does not have contact check
smuX.contact.check
X = SMU channel (a or b)
Function Determine if contact resistance is lower than threshold.
Usage flag = smuX.contact.check() -- check contact resistance against threshold
Remarks Attempting to perform a contact check measurement when any of the following
conditions exists will generate an error:
• Output is off in High-Z mode.
• Current limit set to less than 1mA.
Details See Section 3 for connections.
Also see smuX.contact.threshold, smuX.contact.speed
Example Takes action if contact check on SMU A fails:
if (not smua.contact.check()) then
-- take action
end
Note: 2635/36 does not have contact check
smuX.contact.r
X = SMU channel (a or b)
Function Measure contact resistance.
Usage
rhi, rlo = smuX.contact.r()
rhi – the measured contact resistance on the high/sense high side
rlo – the measured contact resistance on the low/sense low side.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals