2600S-901-01 Rev. C / January 2008 Return to Section Topics 12-101
Series 2600 System SourceMeter® Instruments Reference Manual Section 12: Instrument Control Library
Details See “Status byte and service request (SRQ)” in Appendix D.
Example Reads the service request event register:
serveventreg = status.request_event
print(serveventreg)
Output: 1.29000e+02
The above output indicates that bits B0 (MSS) and B7 (OSB) are set.
status.reset
Function Resets all bits set in the status model.
Usage status.reset()
Remarks
This function clears all status data structure registers (enable, event, NTR and PTR) to
their power up states.