12-102 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 12: Instrument Control Library Series 2600 System SourceMeter® Instruments Reference Manual
status.standard.*
status.standard.condition
status.standard.enable
status.standard.event
* = condition, enable or event
Attribute Standard event register set.
Usage Reads condition, enable and event registers:
standardreg = status.standard.condition
standardreg = status.standard.enable
standardreg = status.standard.event
Writes to enable register:
status.standard.enable = standardreg
Set standardreg to one of the following values:
0 Clears all bits.
status.standard.OPERATION_COMPLETE Sets OPC bit (B0).
status.standard.OPC Sets OPC bit (B0).
status.standard.QUERY_ERROR Sets QYE bit (B2).
status.standard.QYE Sets QYE bit (B2).
status.standard.DEVICE_DEPENDENT_ERROR Sets DDE bit (B3).
status.standard.DDE Sets DDE bit (B3).
status.standard.EXECUTION_ERROR Sets EXE bit (B4).
status.standard.EXE Sets EXE bit (B4).
status.standard.COMMAND_ERROR Sets CME bit (B5).
status.standard.CME Sets CME bit (B5).
status.standard.USER_REQUEST Sets URQ bit (B6).
status.standard.URQ Sets URQ bit (B6).
status.standard.POWER_ON Sets PON bit (B7).
status.standard.PON Sets PON bit (B7).
standardreg can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (OPC), set standardreg to 1 (2
0
).
To set bit B2 (QYE), set standardreg to 4 (2
2
).
To set bit B5 (CME), set standardreg to 32 (2
5
).
To set more than one bit of the register, set standardreg to the sum of their decimal
weights. For example, to set bits B0 and B2, set standardreg to 5 (1 + 4).