Version 1.55 Release - Enhancements Series 3700A Firmware Release Notes
3706A-850SV1.55a / June 2017
4
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
The instrument causes Error 5511 Low-Power when configuring a scan while some channels are
already closed.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
From the web-page, Step by Step will treat scan list of items all from same channel like 1 step
Workaround:
None.
Resolution:
This issue has been corrected.
Enhancements
The new 3760, 3761, 3762, and 3765 switch cards are now supported in this firmware release.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
The instrument should produce a 3-digit exponent in 3706 emulation mode when remotely accessing
print() and printbuffer() commands. Instead it produces a 2-digit exponent like in regular mode.
Workaround:
None.
Resolution:
This issue has been corrected.
Non-critical fixes
None.
PR-59260
PR-44535
RPE-117