ECNEDIFNOCFOERUSAEMRETAERGA
Model 2110
5½ Digit Multimeter
Calibration Manual
2110-905-01 Rev. A / July 2012
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Questions and Answers

  • W
    woodardnicoleAug 12, 2025
    Why does my Keithley Multimeter display '-221 Settings conflict' error?
    • J
      James Armstrong DVMAug 12, 2025
      The error -221 'Settings conflict' on a Keithley Multimeter can occur in several situations. First, it can happen if you send a CONFigure or MEASure command with autorange enabled alongside a fixed resolution. Second, this error appears if you activate math functions and switch to an operation incompatible with the current measurement function. Finally, this error generates when autorange is enabled on the 10A ranges for DCI and ACI.
  • R
    Ryan WeaverAug 22, 2025
    What to do if Keithley 2110-100 Multimeter displays '-213 Init Ignored'?
    • D
      Danielle MartinAug 22, 2025
      The Keithley Multimeter displays the '-213 Init Ignored' error because a measurement is in progress. To solve this, send a device clear command. This will halt the current measurement and return the meter to an idle state.
  • K
    Kevin BaxterAug 31, 2025
    Why does my Keithley 2110-100 Multimeter have erroneous self-test failures?
    • D
      Dennis MckayAug 31, 2025
      Erroneous self-test failures on a Keithley Multimeter may occur due to several reasons. One reason is that signals are present on the input terminal during the self-test procedure. Another reason is that long test leads can act as antennas, picking up AC signals. Also, incorrect power line voltage settings can cause these failures.

Keithley 2110-100 Specifications

General IconGeneral
TypeDigital Multimeter
Display TypeLCD
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ContinuityYes
Diode TestYes
Frequency Accuracy0.01% + 3 digits
Temperature MeasurementYes
InterfaceUSB
Power Supply100 V to 240 V