Section
2: Adjustment Model 2110 5½ Digit Multimeter
Calibration Manual
2-18 2110-905-01 Rev. A / July 2012
Calibration date
1. Update the calibration date by sending the following command:
CAL:PROT:DATE <”MM/DD/YYYY”>
The date <string> must be enclosed in double quotes (“<string>”).
Example: If calibration date is 5 Apr 2012, send the following command:
CAL:PROT:DATE ”04/05/2012”
Calibration due date is automatically calculated for 12 months.
2. Save calibration constants in EEPROM by sending the following command:
CAL:PROT:SAVE
3. Lock calibration by sending the following command:
CAL:PROT:LOCK