In this section: 
Performing accuracy verification .............................................. 3-1 
 
 
Performing accuracy verification 
 
After calibration  is completed, the Model 2110 should  be verified to ensure that its performance is 
within the manufacturer’s specifications.  
 
Perform accuracy verifications using the Fluke 5700A/5725A and 5520A/5522A standards. Refer to 
the following tables for accuracy limits for the specified ranges. 
 
DC voltage accuracy verification 
 
 
DC VOLTAGE ACCURACY VERIFICATION 
Range  Test Point   Accuracy Limits 
    Upper Limit  Lower Limit 
  0.900000 V  0.90001 V  0.89988 V 
10 V  5.0000 V  5.0008 V  4.9992 V 
  9.0000 V  9.0013 V  8.9987 V 
Specifications valid after 2-hour warm-up: 
a. ADC set for continuous trigger operation. 
b. Input bias current < 30 pA at 25 °C. 
c. Input protection 1000 V all ranges (2-W input). 
d. Measurement rate set to 1 PLC. 
e. Repeat the checks above for negative polarity. 
 
Section 3 
Performance verification