Version 1.54 Release - Critical fixes Series 3700A Firmware Release Notes
3706A-850SV1.55a / June 2017
8
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When identical relays are closed/opened in two or more consecutive scan-steps, the Channel Ready
trigger event is generated only when the first identical step is executed. It is not generated when the
subsequent identical steps are executed.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
An error occurs when scan.add() used in a loop for more than 91 times when multiple switching cards
are used.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
Autoranged 4-wire resistance measurement of a 10 ohm resistor returns overflow on the 10 ohm
range when the lead resistance exceeds the Maximum 4WΩ Lead Resistance specification for the
range.
Workaround:
None.
Resolution:
This issue has been corrected. For the case of a 10 ohm resistor with excessive lead resistance, the
autoranging process will now bump the range up to the next highest range (100 ohms) and take the
measurement.
PR56938
PR54842
PR58235

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
BrandKeithley
Model3706A
CategoryMultimeter
LanguageEnglish

Related product manuals