Series 3700A Firmware Release Notes Version 1.52 Release - Non-critical fixes
3706A-850SV1.55a / June 2017
15
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When communicating with the product using USB and using the Read Status Byte bus operation it
always returned 0.
Workaround:
None.
Resolution:
It now returns the correct status byte.
PR47137