Symptom: 
It was observed that reading large amounts of data using VXI would occasionally cause the data to be 
terminated early. 
 
Resolution: 
During heavy amounts of reading the VXI driver would set the EOI bit too early terminating the data 
transfer. This was corrected. 
 
 
Enhancements 
None. 
 
Noncritical fixes 
Models affected: 
3706A, 3706A-S, 3706A-NFP, 3706A-SNFP 
 
Symptom: 
The login window of the Java applet was not sized properly. 
Resolution: 
The resolution of the login window was set to correct values. 
Models affected: 
3706A, 3706A-S, 3706A-NFP, 3706A-SNFP 
 
Symptom: 
channel.calibration.verifydate returns a misleading error for cards that do not support calibration. 
Resolution: 
For cards that do not support calibration the error was changed to “Card not supported”.