EasyManuals Logo
Home>Keithley>AC Power Distribution>6220 DC

Keithley 6220 DC User Manual

Keithley 6220 DC
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
5-44 Delta, Pulse Delta, and Differential Conductance Model 6220/6221 User’s Manual
Return to Section 5 topics
The query form for the arm command (SOUR:DCON:ARM?) is used deter-
mine if Differential Conductance is armed. A returned “1” indicates that Differ-
ential Conductance is armed. A “0” indicates that Differential Conductance is
not armed.
If the Model 6221 is already armed for another action (e.g., Delta or Pulse
Delta), the Differential Conductance arm command will “un-arm” the other
action and arm Differential Conductance.
4. During the arming process, the Model 622x acquires the offset voltage
(V-zero) value of the Model 2182/2182A. The Model 622x factors this value
into the calculation for Differential Conductance voltage.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 DC and is the answer not in the manual?

Keithley 6220 DC Specifications

General IconGeneral
BrandKeithley
Model6220 DC
CategoryAC Power Distribution
LanguageEnglish