4-8 Sweeps Model 6220/6221 User’s Manual
Return to Section 4 topics
Remote sweep operation
Procedures for programming and running a sweep are shown on the following
pages. Each of these procedures includes commands for a typical sweep exam-
ple. Table 4-2 summarizes parameters for each of these examples. See “SCPI
commands — sweeps,” page 4-11 for listings of sweep commands.
Table 4-2
Sweep example parameters
Sweep type Parameters for sweep examples
Linear staircase sweep (page 4-9) Start current: 1mA
Stop current: 10mA
Step current: 1mA
Delay: 1s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 100μA
Compliance: 10V
Custom (list) sweep (page 4-10) # points: 5
Points: 3mA, 1mA, 4mA, 5mA, 2mA
Delay: 3s, 1s, 4s, 5s, 2s
Compliance: 3V, 1V, 4V, 5V, 2V
Source range: auto
Compliance abort: off
Sweep count: 1
Bias current: 50μA