B-8 SCPI Tables (Abridged) Model 6220/6221 User’s Manual
Return to Appendix B tables
Table B-8
System command summary
Command Description Default
:SYST
:COMM Communication interfaces:
:SEL <name> Select interface: SERial, GPIB, or ETHernet.
:GPIB GPIB interface:
:ADDR <NRf> Set primary address: 0 to 30.
:SER RS-232 interface:
:CONT Handshaking:
:RTS <name> Select handshaking: IBFull, RFR, or OFF.
:PACE <name> Set flow control: XON or OFF
:TERM <name> Set output terminator: CR, LF, CRLF, or LFCR.
:BAUD <n> Set baud rate: 300, 600, 1200, 2400, 4800, 9600, 19.2k, 38.4k,
57.6k, or 115.2k.
:SEND <data> Send data via RS-232.
:ENT? Reads data from serial port.
:LOC RS-232 only – Take 622x out of remote.
:REM RS-232 only – Put 622x in remote.
:RWL <b> RS-232 only – Enable or disable local lockout.
:ETH Ethernet interface (6221 only):
:ADDR <string> Set IP address: “n.n.n.n”.
:MASK <string> Set subnet mask: “n.n.n.n”.
:GAT <string> Set Ethernet gateway: ”n.n.n.n”.
:DHCP <b> Enable or disable DHCP.
:SAVE Saves Ethernet setting changes.
:KEY <n> Simulate key press. See Section 13 of Reference Manual.
:KCL <b> Enable or disable key-click. ON
:BEEP Beeper:
:STAT <b> Enable or disable beeper. ON
:PRES Return 622x to PRESET default setup.
:POS <name> Select power-on setup: RST, PRES, SAV0, SAV1, SAV2, SAV3, or
SAV4.
:ERR? Query the latest error code and message.
:CLE Clears error code and message from Error Queue.
:TST System timestamp:
:RES Reset timestamp to zero seconds.
:RNUM System reading number:
:RES Reset reading number to zero.
:VERS? Query revision level of SCPI standard.