35
:ARM[:SEQuence[1]]
:LAYer2
:IMMediate
:COUNt <n>|
INFinite
:COUNt?
:DELay <num>
:DELay?
:SOURce HOLD|
IMMediate|
MANual|BUS|
TLINk|EXTernal
|TIMer
:SOURce?
:SIGNal
:TIMer <num>
:TIMer?
:TCONfigure
:DIRection
SOURce|
ACCeptor
:DIRection?
:ASYNchronous
:ILINe <n>
:ILINe?
:OLINe <n>
:OLINe?
Path to program Arm Layer 2
(scan layer) of scan:
Loop around control source.
Program number of scans (1
to 9999, or INFinite).
Read scan count.
Program delay (0 to
99999.999 sec)
Read delay.
Select Arm Layer 2 control
source.
Read source.
Bypass scan control source
one time only.
Set timer interval (0 to
99999.999 sec).
Read timer interval.
Path to configure Triggers:
Enable (SOUR) or disable
(ACC) Bypass.
Read direction.
Path to configure asyn-
chronous mode:
Select input line (1 to 6).
Read input line.
Select output line (1 to
6).
Read output line.
Table 15: Trigger command summary (Cont.)
Command Description