A-81
APPENDIX A
Scripts
smub.source.rangev = l_sourcev --Set output level
smua.source.limiti = l_cmpl --Set compliance level
smub.source.limiti = l_cmpl --Set compliance level
smua.measure.nplc = 1 --Set measurement aperture
smub.measure.nplc = 1 --Set measurement aperture
smua.measure.autozero = smua.AUTOZERO_AUTO --Set Autozero mode
smub.measure.autozero = smub.AUTOZERO_AUTO --Set Autozero mode
-- Setup SMUA buffer to store all the result(s) in and start testing.
smua.nvbuffer1.clear() --Clear Nonvolatile buffer
smua.nvbuffer1.appendmode = 0 --Append buffer? 0 = No, 1 = Yes
smua.nvbuffer1.collecttimestamps = 0 --Collect Timestamps? 0 = No, 1 = Yes
smua.nvbuffer1.collectsourcevalues = 0 --Collect Source Values? 0 = No, 1 = Yes
smua.source.output = smua.OUTPUT_ON --Enable outputs
smub.source.output = smua.OUTPUT_ON --Enable outputs
smua.source.levelv = l_sourcev -- Program source to level.
smub.source.levelv = l_sourcev -- Program source to level.
smua.measure.count = l_points --Number of points to collect
smua.measure.i(smua.nvbuffer1) -- Measure current and store in reading buffer.
smua.source.output = smua.OUTPUT_OFF
smub.source.output = smub.OUTPUT_OFF
-- Update the front panel display and restore modified settings.
smua.source.levelv = 0
smub.source.levelv = 0
printbuffer(1,l_points,smua.nvbuffer1)
end --function RunHighVoltage(sourcev, points)
--RunHighVoltage(40, 10)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley Series 2600 and is the answer not in the manual?

Keithley Series 2600 Specifications

General IconGeneral
ModelSeries 2600
Measurement FunctionsVoltage, Current, Resistance, Power
ResolutionUp to 6.5 digits
Current Source RangeUp to 10A
Communication InterfacesGPIB, USB
ApplicationsSemiconductor testing
CategorySource Measure Unit (SMU)
ChannelsSingle or Dual Channel (varies by model)