A-48
APPENDIX A
Scripts
Program 10. Transconductance
--[[
Transconductance():
This program sources a voltage bias on a drain-source of a FET (VDS), sources a voltage on
the gate (VGS1), and measures the drain-source current (ID1). Then, another source value
(VGS2) is sourced and the IDS2 is measured.
The transconductance (gfs) is then calculated by taking the change in Ids divided by the
change in VGS.
The drain-source voltage (VDS), Transconductance (gfs), gate-source voltage (VGS), and
drain-source current (ID) are returned.
Required equipment:
(1) Dual-channel Keithley Series 2600 System SourceMeter instrument
(1) SD210 N-channel FET
Running this script creates functions that can be used to create a transconductance test
of FETs. The default values are for an N-channel SD210 FET.
The functions created are:
1. Transconductance(vgsstart, vgsstop, vgssteps, vdsbias)
--Default values vgsstart = 0V, vgsstop = 5V, vgssteps = 100, vdsbias =
10V
2. Check_Comp()
See detailed information listed in individual functions.
To run:
1) From Test Script Builder
- Right-click in the program window, select “Run as TSP”
- At the TSP> prompt in the Instrument Control Panel, type Transconductance()
2) From an external program
- Send the entire program text as a string using standard GPIB Write calls.
Rev1: JAC 6.18.2007
]]--
------------------ Keithley TSP Function ------------------
function Transconductance(vgsstart, vgsstop, vgssteps, vdsbias)--Configure SMUA to source a
user-defined voltage on the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley Series 2600 and is the answer not in the manual?

Keithley Series 2600 Specifications

General IconGeneral
ModelSeries 2600
Measurement FunctionsVoltage, Current, Resistance, Power
ResolutionUp to 6.5 digits
Current Source RangeUp to 10A
Communication InterfacesGPIB, USB
ApplicationsSemiconductor testing
CategorySource Measure Unit (SMU)
ChannelsSingle or Dual Channel (varies by model)

Related product manuals