12-52 Instrument Control Library Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 12 topics
opc function
This function sets the OPC bit in the status register when all overlapped commands are
completed.
Function Sets the Operation Complete status bit when all overlapped commands are completed.
Usage
opc()
Remarks
This function will cause the Operation Complete bit in the Standard Event Status
Register to be set when all previously started local overlapped commands are
complete. Note that each node will independently set their Operation Complete bits
in their own status models.
Any nodes not actively performing overlapped commands will set their bits
immediately. All remaining nodes will set their own bits as they complete their own
overlapped commands.
Details See “Overview” in Appendix D.
Also see waitcomplete
opc