TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011 11
KERN PLJ 720-3AM PLJ 6200-2AM
Product number / type TPLJG 720-3M-A TPLJG 6200-2M-A
Weighing range (Max) 720 g 6200 g
Interval (d) 0.001 g 0.01 g
Reproducibility 0.001 g 0.01 g
Linearity ±0.002 g ±0,02 g
Verification scale interval (e) 10 mg 100 mg
Verification class II II
Minimum weight (Min) 20 mg 50 mg
Settling time (standard) 2 s 2 s
Minimum part weight when
laboratory conditions**
1 mg 10 mg
Minimum part weight when
standard conditions**
10 mg 100 mg
Heating time 4 h 4 h
Adjustment weight internal
Number of reference pieces
when counting the number of
pieces
10, 25, 50, 100, selected freely
Weight units ct, g
Power supply
230V/50Hz AC (Euro), 24V/1A DC
Operating temperature +15°C ... +30°C
Air humidity max. 80% (non-condensing)
Housing (W × D × H) mm
210 x 345 x 155 210 x 345 x 160
Wind breaker [mm]
internal: Ø 150, height 60
external: Ø 160, height 70
Scale plate (stainless steel) Ø 11 cm Ø 16 cm
(Net) weight kg
4.9 kg 5.4 kg
Interface RS-232C
Contamination degree 2
Overvoltage category Category II
Installation height
above sea level
up to 4000 m
Location site Solely indoors

Table of Contents

Other manuals for KERN ALS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALS Series and is the answer not in the manual?

KERN ALS Series Specifications

General IconGeneral
BrandKERN
ModelALS Series
CategoryScales
LanguageEnglish

Related product manuals