TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011 39
8.5 Changing the weight of the internal calibration weight
!
This change can be carried out solely by a specialist holding thorough knowledge of
handling scales.
!
Information on reference weights can be found on the Internet at
http://www.kern-sohn.com.
Display the menu item “Technical adjustment”, see chapter 8.1.
To confirm, press and hold the PRINT button until the sound signal stops.
Press and hold the MENU
button until the sound signal stops. The scale will switch to
the weighing mode again automatically.
Ensure there are no objects on the scale plate. Press the CAL button.
Wait until the blinking weight value of the adjustment weight is displayed.
During
blinking, place the required adjustment weight carefully in the middle of the
scale plate. The blinking symbol will disappear.
Wait until the stabilization indicator is displayed, remove the adjustment weight.
Press and hold the PRINT button until the sound signal stops.
The weight of the internal adjustment weight will be changed.
After the successful change, the scale will switch to the weighing mode again
automatically.

Table of Contents

Other manuals for KERN ALS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALS Series and is the answer not in the manual?

KERN ALS Series Specifications

General IconGeneral
BrandKERN
ModelALS Series
CategoryScales
LanguageEnglish

Related product manuals