CKE-BA-e-2132 29
Place the adjustment weight and confirm, pressing TARE. The <wait> and
<finish> symbols will be displayed consecutively.
After the successful adjustment, the scales will switch to the weighing mode again
automatically.
If an adjustment error occurs (e.g. there are any items on the scales plate), the dis-
play will show <wrong>. Switch the scales off and repeat the adjustment process.
11.2.4 Taking over the placed weight as the PRE-TARE value <Ptare actuel>
Place the scales vessel.
Display the <Ptare> menu item and confirm, pressing TARE.
To take over the weight of the placed weight as the PRE-TARE value, choose
the <actuel> option, using navigation buttons .
Confirm, pressing TARE. The <wait> symbol will be displayed.
The scales vessel weight will be saved as tara.
Remove the scales vessel, the following will be displayed: (TARE) symbol and
tara weight with a negative symbol.
Place a filled scales vessel.
Wait until the stabilization indicator is displayed ( ).
Read out the net weight.
The entered tara value will be used until you enter a
new one. To delete it,
press TARE or confirm the menu item <clear>, pressing TARE.

Table of Contents

Other manuals for KERN CDS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CDS Series and is the answer not in the manual?

KERN CDS Series Specifications

General IconGeneral
BrandKERN
ModelCDS Series
CategoryScales
LanguageEnglish

Related product manuals