15 Help for any minor failures
If there are any programme execution problems, the scales should be switched off
and disconnected from the mains for a while. Next, the weighing process should be
started anew.
Problem Possible cause
The weight indicator
is not lit
The scales is not on.
Interrupted mains connection (mains cable not
connected/damaged).
Mains voltage failure.
The weight indication
keeps fluctuating
Draft/air movements.
Table/air vibrations.
The scales plate is in contact with foreign bodies.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the interfer-
ing device).
The weighing result
is clearly wrong
The scales indication was not reset.
Incorrect adjustment.
Scales not placed on a level surface.
There are heavy temperature fluctuations.
The heating time not observed.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the interfer-
ing device).