Page #32 background imageLoading...
Page #32 background image
Settings in the menuFTC-BA-e-0720
32User manual ME-Nr.22017145 01/07
4.4 Application settings (APPLICATION)
4.4.1 COUNT – settings for counting
VAr-SPL Adaptation of the reference quantity
ON
The reference quantity can be changed in operating mode
OFF
Counting only with defined reference quantities
SPL-qtY Reference quantity
Sq1 ... Sq5
Define 5 fixed reference quantities
Min.reFW Monitoring the minimum reference weight
OFF
No monitoring of the minimum reference weight
97.5, 99.0,
99.5
Monitoring the minimum reference weight so that a counting accuracy of
97.5 %, 99.0 % or 99.5 % is achieved
rEF.OPt Optimizing the average piece weight
OFF
No reference optimization
AUtO
Automatic reference optimization